Configuration
3.2.5 Alarms
Use the following steps to configure the alarms, which are located in the Resource Function
Block:
1. Set the resource block to OOS.
2. Set WRITE_PRI to the appropriate alarm level (WRITE_PRI has a selectable range of
priorities from 0 to 15, see Alarm Priority Levels” on page 31. Set the other block alarm
parameters at this time.
3. Set CONFIRM_TIME to the time, in
1
/32 of a millisecond, that the device will wait for
confirmation of receiving a report before trying again (the device does not retry if
CONFIRM_TIME is 0).
4. Set LIM_NOTIFY to a value between zero and MAX_NOTIFY. LIM_NOTIFY is the
maximum number of alert reports allowed before the operator needs to acknowledge
an alarm condition.
5. Enable the reports bit in FEATURES_SEL. (When Multi-bit alerts is enabled, every active
alarm is visible for any of the eight sensors, generated by a PlantWeb
®
and Field
Diagnostics alert. This is different than only viewing the highest priority alarm.)
6. Set the resource block to AUTO.
For modifying alarms on individual function blocks (AI or ISEL blocks), refer to Appendix D:
Function Blocks.
3.2.6 Damping
Use the following steps to configure the damping, which is located in the transducer function
block:
1. Set Sensor Mode to Out of Service.
2. Change DAMPING to the desired filter rate (0.0 to 32.0 seconds).
3. Set Sensor Mode to In Service.
3.2.7 Configure the differential sensors
Use the following steps to configure the differential sensors:
1. Set Dual Sensor Mode to Out of Service.
2. Set Input A and Input B to the sensor values that are to be used in the differential
equation diff = A–B. (Note: Unit types must be the same.)
3. Set the DUAL_SENSOR_CALC to either Not Used, Absolute, or INPUT A minus INPUT B.
4. Set Dual Sensor Mode to In Service.